Description
STENT
| Stent Material | Cobalt Chromium (L605) |
| Strut Thickness | 65µm (0.065mm or 0.026″) |
| Stent Diameter (mm) | 2.00, 2.25, 2.50, 2.75, 3.00, 3.50, 4.00, 4.50 |
| Stent Length (mm) | 8, 13, 16, 19, 24, 29, 32, 37, 40, 44, 48 |
| Mean Recoil | 3.38% |
| Mean Foreshortening | 0.29% |
DRUG / POLYMER
| Drug | Sirolimus |
| Equivalent Drug Dose | 1.25 µg/ mm² |
| Polymer | Biodegradable & Biocompatible (PLLA + PLGA) |
DELIVERY SYSTEM
| Delivery System | Rapid Exchange |
| Nominal Pressure | 9 atm (Refer IFU) |
| Rated Burst Pressure | Rated Burst Pressure: 14/16 atm* |
*RBP 14 atm for 4.50 mm diameter, all lengths.
*RBP 14 atm for all diameter with length >40mm. (Refer IFU for more details)
| Balloon Overhang | ≤ 0.5 mm |
| Shaft Outer Diameter | Proximal: 2.13 F, Distal: 2.7 F (Refer IFU for more details) |
| Radiopaque Markers | Two Platinum / Iridium Markers |
| Usable Catheter Length | 142 cm (Refer IFU for more details) |
| Guide Catheter Compatibility | 5F (Min. I. D. 0.056″ / 1.42 mm) |
| Tip Entry Profile | 0.016″ |